Products
SKAN_Berner_claire_neo_Sicherheitswerkbank
Only available in Switzerland

Requirements

  • Optimum personal, product and contamination protection
  • Aseptic production of cytostatic drugs, parenteral drugs
  • Handling of microbiological agents of classes S1, S2, S3 and S4

Tested

Acc. to DIN 12980, DIN EN 12469 and NSF 49
  • Solution

    The new generation of safety workbenches provides protection at the highest level. Proven technologies are combined with scientifically developed solutions, resulting in perfect protective equipment. With the Claire Class II safety workbench, personal, product and contamination protection are given the highest priority.

  • Functional principle

    As opposed to the common 2-filter technology, the third filter stage not only offers a significantly higher safety level due to its innovative filter arrangement, but also numerous other advantages: Redundant HEPA filter system with excellent overall separation levels. The main filter stage immediately below the work surface separates particulate contaminants directly via an extremely short air channel. This prevents unnecessary contamination of inaccessible areas. Essential air-ducting components such as filters, fans and air ducts remain free of contamination.

  • Characteristics
    • Protection of personnel and products at the highest level
    • Optimal ergonomics
    • HEPA cartridge filter, Low noise level
    • The latest generation touch display
    • GreenTec reduces energy operating costs
    • Optimal legroom even with 3-filter units
    • Detection system for people movement
    • Screen and microscope workstations
    • Extensive accessories
  • Field of application
    • Laboratories and Research
    • Biotec and Life Sciences
    • Hospital and Pharmacy
    • Medical Technology
    • Food Industry
    • Customized solutions
  • Additional products and options
    • Electrically height-adjustable base stand
    • UVC light
    • Utility supply connections and particle monitoring
    • Interfaces (e.g. USB) and CH sockets
    • Monitor-based and microscope workstations
    • Weighing work plate
    • FlexDuct exit air connection
    • Further options upon request

SKAN Pure Solutions

Technical specification

Technical Data (in mm)Claire Pro 130Claire Pro 160Claire Pro 190
Outer dimensions (W x D x H)1352 x 815 x 2008-22771654 x 815 x 2008-22771957 x 815 x 2008-2277
Inside dimensions, working space (W x D x H)1257 x 640-700 x 6001559 x 640-700 x 6001862 x 640-700 x 600
Power consumption (GMP/ECO)280 / 190 WNA530 / 335 W

Recommendations


Related services